About capacitive level sensor
We are measuring 30 wt% KOH in the pipe, and the wall thickness is somewhere between 2.87 mm to 4.55 mm (depending on 3/4” vs. 1” and schedule 40 vs. 80). 30 wt% KOH has a conductivity of~1.2-1.3 S/cm. The sensor I’m interested in is SKU: ATO-CLS-CE15. Can it be used?
From:
Alden
|
Date:
27/09/2024
Was this review helpful?
Yes
No
(0/0)
Sorry, this unit SKU: ATO-CLS-CE15 has a detection distance of only 2mm. If the wall thickness reaches 2.87 mm to 4.55 mm, the sensor can Not measure properly. We recommend the square capacitive proximity sensor, SKU: ATO-SPS-C2010, with a detection distance of 1-10mm. Please check if it meets your requirements, see spec on: https://www.ato.com/square-proximity-sensor-capacitive